Precautions For Use Of Inoculation Loop Sterilizer

Mar 29, 2022Leave a message


Due to the characteristics of high temperature and high risk, the following points must be paid attention to in use to avoid accidents:


1: There must be a good ground wire in the power socket to prevent electric shock.


2: Hc-100a and Hc-100b have two sterilization temperatures, the high temperature range is 850°C, and the low temperature range is 500°C. If the sterilizer is not used for a short time, the sterilizer can be set to the low temperature range, and the equipment is in the preheating state. When sterilization is required, adjust to the high temperature level, wait for 5 minutes, and then sterilize immediately; if it will not be used for a long time, turn off the power switch.


3: The sterilizer must be used in a dry environment. Cannot be used when the ambient humidity is greater than or equal to 95%RH


Bacterial instrument. When using the sterilizer, there should be no water droplets on the user's hands.


4: It is forbidden to scratch the infrared ceramic tube with sharp objects, because there is a heating wire in the ceramic tube to prevent the inner wall of the ceramic tube


The exposed wires leak after being scratched.


5: It is forbidden to extend small objects such as inoculation rings into the cooling holes or connections of the equipment.


6: Regularly check the inner surface of the heater for obvious scratches or residue accumulation.


7: If the heater is found to be defective, it should be replaced as soon as possible.


8: The sterilization time of the inoculation loop/needle and other items in the heating chamber is 5-8 seconds, and the burning dwell time should not be extended to avoid the handle


Some heat up, reducing the service life.


9: This product can also be used for disinfection of microbial culture tubes with glass material less than φ14mm, but there is no liquid in the tube


body and substances to avoid liquid splashes and personal injury or explosion accidents in them. We cannot use plastics and other materials that are not resistant to high temperatures


Put them in the heating tube.


10: In an emergency, the power switch can be turned off directly to terminate the operation.


11: During operation, the peripheral temperature of the heater will also be higher. Please avoid contact and burning with flammable materials.


12: When cleaning and maintenance, it is recommended to turn off the power of the host, unplug the power cable, keep it cool for more than 1 hour, without electricity


operate.


13: When cleaning, wipe the outside of the instrument with a cloth dampened with isopropyl alcohol or other mild detergent. caution,


When cleaning and wiping the heater periphery and the connector of the power connection wire, do not drip the cloth to avoid short circuit


Instrument road damage.